Nimbus™ II Flex infusion pump
Nimbus™ II Flex is a sleek and cutting-edge ambulatory infusion pump. It affords the sophistication of a high-end electronic pump designed to compete in the disposable market. Setting up can be as easy as rate, volume and run, Nimbus™ II Flex allows you to monitor infusions on the go.


Demonstration Videos
Guidelines, Specifications & User Manuals
Helpful Information
NIMBUS™ II FLEX PUMP PRODUCT SPECIFICATIONS
Product Dimension 4.25 x 2.28 x 1.5 (in), 108 x 58 x 40 (mm)
Product Weight 6.14 oz, 174g
Pump Mechanism Linear Peristaltic
Operating Mode Continuous mode
Max. Infusion Pressure 30 psi
Max. Time to Occlusion Alarm 20 minutes
VTBI 1 – 1500mL by 0.1mL increment (single infusion VTBI maximum
value is limited by remaining pump lifetime VTBI)
Rate 0.1 mL/h – 135 mL/h in 0.1mL/h increments
Flow Rate Accuracy +/- 5%
Power Supply AC/DC via Micro USB port, internally powered by battery (battery
not rechargeable)
Battery Life 400 h or 1500mL of volume infused
Alarms • Cassettes loading error
- System error
- Firmware error
- Upstream occlusion
- Downstream occlusion
- Battery depleted
- Invalid infusion parameter
- Pump unattended
- Infusion complete
- Power off and replace battery
- Due for service
Warnings Battery low
Adminstration Set Options • Bag spike, luer lock
- Set-based free flow
protection
- Slide clamp
- 1.2 micron filter (air eliminating)
- 0.2 micron filter (air eliminating)
- Luer lock adaptor
Optional Accessories Carrying pouches for pump, Drug bag, Hardshell
Standards Electrical Safety: IEC 60601-1
Pump Performance: IEC 60601-2-24
Electromagnetic Compatibility: IEC 60601-1-2
Usability and Human Factor: IEC 60601-1-6 ed3/ IEC 62366
Alarm System: IEC 60601-1-8
Home Healthcare Environment: IEC 60601-1-11
Similar Instructions