Nimbus™ II Flex

Nimbus™ II Flex is a sleek and cutting-edge ambulatory infusion pump. It affords the sophistication of a high-end electronic pump designed to compete in the disposable market. Setting up can be as easy as rate, volume and run, Nimbus™ II Flex allows you to monitor infusions on the go.

Demonstration Videos

Guidelines, Specifications & User Manuals

Brochure

Quick Reference

Patient Guide

Accessories Catalog

Helpful Information

NIMBUS™ II FLEX PUMP PRODUCT SPECIFICATIONS

Product Dimension 4.25 x 2.28 x 1.5 (in), 108 x 58 x 40 (mm)

Product Weight 6.14 oz, 174g

Pump Mechanism Linear Peristaltic

Operating Mode Continuous mode

Max. Infusion Pressure 30 psi

Max. Time to Occlusion Alarm 20 minutes

VTBI 1 – 1500mL by 0.1mL increment (single infusion VTBI maximum

value is limited by remaining pump lifetime VTBI)

Rate 0.1 mL/h – 135 mL/h in 0.1mL/h increments

Flow Rate Accuracy +/- 5%

Power Supply AC/DC via Micro USB port, internally powered by battery (battery

not rechargeable)

Battery Life 400 h or 1500mL of volume infused

Alarms • Cassettes loading error

  • System error
  • Firmware error
  • Upstream occlusion
  • Downstream occlusion
  • Battery depleted
  • Invalid infusion parameter
  • Pump unattended
  • Infusion complete
  • Power off and replace battery
  • Due for service

Warnings Battery low

Adminstration Set Options • Bag spike, luer lock

  • Set-based free flow

protection

  • Slide clamp
  • 1.2 micron filter (air eliminating)
  • 0.2 micron filter (air eliminating)
  • Luer lock adaptor

Optional Accessories Carrying pouches for pump, Drug bag, Hardshell

Standards Electrical Safety: IEC 60601-1

Pump Performance: IEC 60601-2-24

Electromagnetic Compatibility: IEC 60601-1-2

Usability and Human Factor: IEC 60601-1-6 ed3/ IEC 62366

Alarm System: IEC 60601-1-8

Home Healthcare Environment: IEC 60601-1-11

Related Devices